FiD Gas chromatograph Introduction

The GC112N FID gas chromatograph is a microcomputer based, high-performance, low-priced, and newly designed universal gas chromatograph. It has the advantages of high stability and reliability, simple and reasonable structure, convenient operation, and beautiful appearance. This instrument has a wide range of applications and is suitable for trace detection of pollution in environmental protection, atmosphere, water sources, etc; Analysis, monitoring, and research of toxins; biochemistry; Clinical application; Pathological and viral research; Food fermentation: petrochemical industry; Petroleum processing; Oil analysis; Geological and mineral exploration research; Organic chemistry; Synthesis research; health quarantine; Analysis and research on public hazard detection. The main components of GC112N gas chromatograph

GC112 FID gas chromatograph is equiped with a FID detector.FID is the abbreviation of Flame flame ionization detector which is a high sensitivity universal detector.
It has response to almost all organic substances, but has no response or very little response to inorganic compound substances, inert gases or undivided substances in the flame.

FID gas chromatograph need use pure hydrogen as flame gas.pure nitrogen as carrier gas.you can consider onsite hydrogen generator and nitrogen generator to replace dangerous gas cylinder.

Features of FID gas chromatograph

Features & Specification
Feature:
1.Standard PC side reverse control software
2.Host with 7-inch color touch screen
3.Gas shortage alarm protection function
4.Split flow/split ratio can be automatically controlled to save carrier gas.
5.Configure automatic sampler installation and positioning interface 
6.The multi-core, 32-bit embedded hardware system 
7.Using logarithmic amplifier, detection signal no cut-off value, good peak shape, extensible synchronous external trigger function
8.One-button start function
9.It has perfect system self-check function and fault automatic identification function.
10.With 8 external event extension function interface
11.RS232 communication port and LAM network port

fid gas chromatograph
fid gas chromatograph

Technical Specifications:

Column temperature box
Content product22L
Temperature control range5ºC ~ 400ºC at room temperature
Temperature control accuracy±0.1ºC
Heating rate0.1 ~ 60ºC / min
Program temperature rise order9
Program heating repeatability≤ 2%
Cooling wayopen the door after
Cooling speed≤10 mins (250ºC ~ 50ºC)
Control software function(GC112N only)
Column temperature box controlDetector controlInjector controlMap display
Sampler
Temperature control range7ºC ~ 420ºC at room temperature
Temperature control methodindependent temperature control
Carrier gas flow control modeconstant pressure
Number of simultaneous installations3 at most
Type of injection unitfilling column, shunt
Split ratiosplit ratio display
Cylinder pressure range0 ~ 400kPa
Cylinder pressure control accuracy0.1kPa
Flow setting rangeH2 0 ~ 200ml / minN2 0 ~ 150ml / min
Detector
FID, TCD optional
Temperature controlMax. 420ºC
Number of simultaneous installations2 at most
Ignition functionautomatic
Hydrogen flame ionization detector (FID)
Detection limit≤ 3×10 g/s (n-hexadecane)
Baseline noise≤ 5× 10-14A
Baseline drift≤ 6× 10-13A
Dynamic range107
RSD3% or less
Thermal conductivity detector (TCD)
Sensitivity5000mV•mL/mg (n-cetane)
Baseline noise≤ 0.05 mV
Baseline drift≤ 0.15mV / 30min
Dynamic range105
Supply voltageAC220V±22V, 50Hz±0.5Hz
Power3000W
Please kindly email us to [email protected] if you want to learn more specification

FID gas chromatograph need pure hydrogen as flame gas.below is the link of PEM pure hydrogen generator.

FID gas chromatograph also need dry compressed air as or zero air as Auxiliary combustion gas.below is the related link